Table 1 Results of replicate measurements of NIST SRM 987 in three different mass spectrometric sessions.
Session | Reference date | 88Sr (V) | 90Sr1,2 (cps) | Average darknoise (cps) | 90Sr/88Sr1,2 Abundance sensitivity | 3SD detection limit | Number of analysis |
|---|---|---|---|---|---|---|---|
Bremen | January 22nd, 2020 | 20.3 | 0.0118 (29) | 0.0070 | 9.3 (3.0) × 10−12 | 3.7 × 10−12 | 4 |
Fukushima-1 | April 3rd, 2021 | 27.2 | 0.0168 (34) | 0.0125 | 1.0 (2.2) × 10−11 | 3.1 × 10−12 | 13 |
Fukushima-2 | June 5th, 2021 | 26.3 | 0.0136 (29) | 0.0096 | 8.3 (1.8) × 10−12 | 2.7 × 10−12 | 17 |