Table 1 Image specifications.

From: Dimensions and forms of artefacts in 1.5 T and 3 T MRI caused by cochlear implants

 

T2-TSE

T2-TIRM

3D-MPRAGE

DTI

TR (MS)

5800

8000

7.8

7000

TE (MS)

95

87

2.94

95

TI (MS)

2370

1100

Flip angle

90°/150°

90°/150°

90

Echo train length

16

16

1

64

Pixel bandwidth (Hz/Px)

220

250

250

1500

Acquisition matrix

218 × 384

175 × 320

187 × 256

64 × 128

Reconstructed matrix

312 × 384

250 × 320

208 × 256

256 × 256

Field of view (mm)

178 × 220

171 × 220

178 × 220

220 × 220

Slice thickness (mm)

2

3

1

3

Gap between slices (mm)

0

0.6

0

0,3

Number of slices

75

45

176

43

Number of concatinations

3

2

1

1

Parallel imaging mode

GRAPPA

GRAPPA

GRAPPA

Parallel imaging factor

2

2

2

Number of diffusion encoding directions

20

B-value

1000

Voxel size (mm)

0.573 × 0.573 × 2

0.688 × 0.688 × 3

0.859 × 0.859 × 1

0.859 × 0.859 × 3

Acquisition time

3 × 54.40 s

2 × 1 min 20 s

3 min 36 s

2 min 50 s

  1. TR (ms): Repetition time (in milliseconds), TE: Echo time, TI: Inversion time, Hz/Px: Herz per pixel, mm: millimeter.