Table 1 Saturation power (\(P_{sat}\)), detection efficiency (\(\eta \)) and average background count rate (Bkg) before and after optimization with different methods.

From: Dynamic compensation of stray electric fields in an ion trap using machine learning and adaptive algorithm

 

\(P_{sat}\) (\(\mu W\))

\(\eta \) (%)

\({\bar{Bkg}}(1/s)\)

Initial

4.11(1)

0.971(2)

1184(34)

ADAM-45

1.86(1)

0.938(1)

1227(35)

MLOOP-45

1.86(1)

0.989(1)

1291(35)

  1. \(P_{sat}\) and \(\eta \) values are calculated by fitting the theoretical formula for Fluorescence Rate versus Laser Power to the experimental data29,30,31. A clear drop in saturation power is observed after ADAM and MLOOP optimization of individual electrodes indicating a reduction of the micromotion.