Table 4 Comparison of diagnostic abilities of PhNRs and OCT in optic nerve disorders.
From: Photopic negative response recorded with RETeval system in eyes with optic nerve disorders
| Ā | Overall | Controlā+āAcute ON/ION | Controlā+āChronic ON/ION | Controlā+āTON/DOA | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
AUC | Sensitivity (%) | Specificity (%) | AUC | Sensitivity (%) | Specificity (%) | AUC | Sensitivity (%) | Specificity (%) | AUC | Sensitivity (%) | Specificity (%) | |
P72 | 0.790 | 66.7 | 77.8 | 0.735 | 76.9 | 61.1 | 0.772 | 73.9 | 66.7 | 0.844 | 75.0 | 88.9 |
Pmin | 0.784 | 72.9 | 72.2 | 0.689 | 61.5 | 77.8 | 0.786 | 73.9 | 69.4 | 0.841 | 75.0 | 88.9 |
Pratio | 0.833 | 77.1 | 83.3 | 0.784 | 69.2 | 83.3 | 0.833 | 78.3 | 80.6 | 0.833 | 83.3 | 83.3 |
Wratio | 0.857 | 75.0 | 83.3 | 0.810 | 76.9 | 75.0 | 0.877 | 82.6 | 80.6 | 0.829 | 83.3 | 72.2 |
cpRNFLT | 0.764 | 75.0 | 91.7 | 0.789 | 69.2 | 100 | 0.935 | 95.7 | 88.9 | 0.995 | 100 | 97.2 |
GCCT | 0.853 | 81.3 | 80.1 | 0.567 | 53.9 | 69.4 | 0.920 | 91.3 | 91.7 | 0.998 | 100 | 97.2 |