Table 9 Optimal threshold of each method as determined by Youden’s J Statistics with 95% confidence interval.

From: Evaluating diagnostic accuracy and determining optimal diagnostic thresholds of different approaches to [68Ga]-DOTATATE PET/MRI analysis in patients with meningioma

Methods

J

Optimal Threshold

Sensitivity

Specificity

SUV

75.27

 > 4.7

77.7% (70.8–83.4)

97.6% (87.4–99.9)

SUVRsss

69.07

 > 3.2

86.1% (80.1–90.6)

82.9% (68.7–91.5)

SUVRpituitary

67.32

 > 0.3

79.5% (72.7–85.0)

87.8% (74.5–94.7)

SUVRnorm

51.34

 > 62.6

80.6% (73.9–85.9)

70.7% (55.5–82.4)