Table 1 Intraclass correlation (ICC) between the five single measurements of each device for the investigated parameters.

From: Comparison of three validated systems to analyse spinal shape and motion

 

Lordosis (95% CI)

Kyphosis (95% CI)

RoF (95% CI)

RoE (95% CI)

RoM (95% CI)

MediMouse

0.986 (0.976–0.993)

0.986 (0.976–0.993)

0.968 (0.946–0.983)

0.951 (0.916–0.974)

0.982 (0.969–0.991)

Raster stereography

0.989 (0.982–0.994)

0.968 (0.949–0.983)

Epionics Spine

0.988 (0.979–0.994)

0.999 (0.999–1.000)

0.988 (0.979–0.994)

0.993 (0.988–0.997)

  1. CI confidence interval, RoM range of motion, RoF range of flexion, RoE range of extension.