Table 1 Values of \(<{\bar{y}}_F>\), \(\sigma\) \(_{<{\bar{y}}_F\!>}\), \(<{\bar{y}}_D>\), and \(\sigma\) \(_{<{\bar{y}}_D\!>}\) obtained with both systems for four representative energies.
From: Microdosimetry performance of the first multi-arrays of 3D-cylindrical microdetectors
\(E_{beam}\) | \(<{\bar{y}}_F>\) \(\pm \sigma\) \(_{<{\bar{y}}_F\!>}\) (keV/\(\mu m\)) | \(<{\bar{y}}_D>\) \(\pm \sigma\) \(_{<{\bar{y}}_D\!>}\) (keV/\(\mu m\)) | ||||
|---|---|---|---|---|---|---|
(MeV) | Pixel-type | Pad-type | Difference | Pixel-type | Pad-Type | Difference |
6.0 | 16.40 ± 0.06 | 16.30 ± 0.04 | 0.6 % | 17.25 ± 0.04 | 17.56 ± 0.07 | 1.8 % |
8.0 | 11.31 ± 0.12 | 11.46 ± 0.05 | 1.3 % | 12.25 ± 0.14 | 12.06 ± 0.05 | 1.6 % |
10.0 | 8.88 ± 0.20 | 9.20 ± 0.12 | 3.5 % | 9.45 ± 0.20 | 9.94 ± 0.22 | 5.1 % |
18.0 | 5.12 ± 0.12 | 5.65 ± 0.04 | 9.8 % | 5.59 ± 0.13 | 6.06 ± 0.07 | 8.1 % |