Table 1 Values of \(<{\bar{y}}_F>\), \(\sigma\) \(_{<{\bar{y}}_F\!>}\), \(<{\bar{y}}_D>\), and \(\sigma\) \(_{<{\bar{y}}_D\!>}\) obtained with both systems for four representative energies.

From: Microdosimetry performance of the first multi-arrays of 3D-cylindrical microdetectors

\(E_{beam}\)

\(<{\bar{y}}_F>\) \(\pm \sigma\) \(_{<{\bar{y}}_F\!>}\) (keV/\(\mu m\))

\(<{\bar{y}}_D>\) \(\pm \sigma\) \(_{<{\bar{y}}_D\!>}\) (keV/\(\mu m\))

(MeV)

Pixel-type

Pad-type

Difference

Pixel-type

Pad-Type

Difference

6.0

16.40 ± 0.06

16.30 ± 0.04

0.6 %

17.25 ± 0.04

17.56 ± 0.07

1.8 %

8.0

11.31 ± 0.12

11.46 ± 0.05

1.3 %

12.25 ± 0.14

12.06 ± 0.05

1.6 %

10.0

8.88 ± 0.20

9.20 ± 0.12

3.5 %

9.45 ± 0.20

9.94 ± 0.22

5.1 %

18.0

5.12 ± 0.12

5.65 ± 0.04

9.8 %

5.59 ± 0.13

6.06 ± 0.07

8.1 %