Table 4 Discriminating ability of the regular normative database on macular ganglion cell complex thickness for glaucoma detection.
Glaucoma (n = 185) | Normal (n = 102) | |
|---|---|---|
Regular normative database | ||
Superior/inferior | ||
Abnormal thinning | 159 | 27 |
Within normal range | 26 | 75 |
Sensitivity (95% CI) | 0.859 (0.801–0.906) | |
Specificity (95% CI) | 0.735 (0.639–0.818) | |
Positive likelihood ratio | 3.25 (2.34–4.51) | |
Negative likelihood ratio | 0.19 (0.13–0.28) | |
Positive predictive value | 0.855 (0.809–0.891) | |
Negative predictive value | 0.743 (0.665–0.808) | |
Accuracy | 0.815 (0.766–0.859) | |
Regular normative database | ||
GChart | ||
Abnormal thinning | 169 | 26 |
Within normal range | 16 | 76 |
Sensitivity (95% CI) | 0.914 (0.863–0.950) | |
Specificity (95% CI) | 0.745 (0.649–0.826) | |
Positive likelihood ratio | 3.58 (2.56–5.01) | |
Negative likelihood ratio | 0.12 (0.07–0.19) | |
Positive predictive value | 0.867 (0.823–0.901) | |
Negative predictive value | 0.826 (0.746–0.885) | |
Accuracy | 0.854 (0.807–0.893) | |