Table 1 Parameters set on the PinAAcle 900H flame/furnace Atomic Absorption Spectrophotometer (AAS).
Metal | Lamp specifications | Flame type | |||
|---|---|---|---|---|---|
Band width (nm) | Wavelength (nm) | Lamp current (mA) | Detection limits (ppm) | ||
Cu | 0.7 | 324.75 | 15–25 | 0.001 | Air |
Co | 0.2 | 240.73 | 15–25 | 0.006 | Air |
Fe | 0.2 | 248.33 | 12–25 | 0.002 | Air |
Mn | 0.7 | 279.48 | 12–25 | 0.002 | Air |
Pb | 0.7 | 283.31 | 10–25 | 0.01 | Air |
Zn | 0.7 | 213.86 | 8–25 | 0.002 | Air |