Table 8 Comparison of detection limits using this proposed technique and other conventional methods.
Analytical technique | LODs | Ref. | |
|---|---|---|---|
Pt | Pd | ||
Pb-FA Spark-OES | 10 ng g−1 | 100 ng g−1 | 25 |
Pb-FA LA-ICP-MS | 30 ng g−1 | 25 ng g−1 | |
Pb-FA GD-MS | 7 ng g−1 | 9 ng g−1 | |
NiS-FA LA-ICP-MS (dynamic reaction cell) | 20 ng g−1 | 28 ng g−1 | 26 |
NiS-FA LA-ICP-MS (focusing sector field MS) | 11 ng g−1 | 17 ng g−1 | 27 |
Pb-FA femtosecond LA-ICP-MS | 6 ng g−1 | 9 ng g−1 | 24 |
Pb-FA LA-ICP-MS | 0.06 ng g−1 | 0.03 ng g−1 | 29 |
Bi-FA LA-ICP-MS | 0.074 ng g−1 | 0.037 ng g−1 | This work |