Table 8 Comparison of detection limits using this proposed technique and other conventional methods.

From: Bismuth fire assay preconcentration and empirical coefficient LA-ICP-MS for the determination of ultra-trace Pt and Pd in geochemical samples

Analytical technique

LODs

Ref.

Pt

Pd

Pb-FA Spark-OES

10 ng g−1

100 ng g−1

25

Pb-FA LA-ICP-MS

30 ng g−1

25 ng g−1

Pb-FA GD-MS

7 ng g−1

9 ng g−1

NiS-FA LA-ICP-MS (dynamic reaction cell)

20 ng g−1

28 ng g−1

26

NiS-FA LA-ICP-MS (focusing sector field MS)

11 ng g−1

17 ng g−1

27

Pb-FA femtosecond LA-ICP-MS

6 ng g−1

9 ng g−1

24

Pb-FA LA-ICP-MS

0.06 ng g−1

0.03 ng g−1

29

Bi-FA LA-ICP-MS

0.074 ng g−1

0.037 ng g−1

This work