Table 3 Subjective comparison of TFI and FBP at the same tube current–time products: CT findings.
Tube current time products (mAs) | Score | |||||
|---|---|---|---|---|---|---|
10 | 25 | 50 | 75 | 100 | ||
Noise | TFI | 1.2 ± 0.4* | 1.7 ± 0.5* | 2.0 ± 0.0* | 2.3 ± 0.4* | 2.0 ± 0.0* |
FBP | 4.8 ± 0.4* | 4.5 ± 0.5* | 3.8 ± 0.4* | 3.5 ± 0.5* | 3.6 ± 0.5* | |
Streak artifact | TFI | 3.0 ± 0.0 | 2.3 ± 0.5* | 2.1 ± 0.3* | 1.8 ± 0.4* | 2.1 ± 0.3* |
FBP | 3.0 ± 0.0 | 3.5 ± 0.5* | 4.3 ± 0.4* | 3.7 ± 0.5* | 3.8 ± 0.4* | |
Overall image quality | TFI | 4.2 ± 0.4* | 4.0 ± 0.0* | 4.0 ± 0.0* | 4.0 ± 0.0* | 4.0 ± 0.0* |
FBP | 1.8 ± 0.4* | 1.8 ± 0.4* | 2.0 ± 0.0* | 2.5 ± 0.5* | 2.4 ± 0.5* | |