Table 2 Comparison of the CEA concentration as detected by the developed prototype and ELISA.

From: Simple diagnosis of cancer by detecting CEA and CYFRA 21-1 in saliva using electronic sensors

Sample no.

Concentration predicted by ELISA (ng/mL)

Concentration predicted by developed prototype (ng/mL)

OSCC/healthy sample

1

250

250

OSCC sample

2

102

100–400

OSCC sample

3

22.17

20–100

OSCC sample

4

6.47

5–10

OSCC sample

5

13.1

10–20

OSCC sample

6

10.685

10–20

OSCC sample

7

19.04

10–20

OSCC sample

8

150.81

150

OSCC sample

9

201.6

200

OSCC sample

10

2

1–5

Healthy sample

11

1.5

1

Healthy sample

12

0.0836

 < 0.001

Healthy sample

13

0.0887

 < 0.001

Healthy sample

14

0.0578

 < 0.001

Healthy sample