Table 3 Comparison of qualitative indices of image quality and tumor assessment between ssEPI DWI and msEPI DWI.

From: Comparison of single-shot EPI and multi-shot EPI in prostate DWI at 3.0 T

 

Reader 1 (9 years of experience)

Reader 2 (13 years of experience)

Reader 2 (23 years of experience)

ssEPI

msEPI

p value

ssEPI

msEPI

p value

ssEPI

msEPI

p value

Anatomical distortion

3.30 ± 0.80

3.67 ± 0.56

 < 0.001

3.05 ± 0.78

3.25 ± 0.74

 < 0.001

3.01 ± 0.97

3.63 ± 0.58

 < 0.001

Prostate edge clarity

2.98 ± 0.48

3.73 ± 0.48

 < 0.001

3.04 ± 0.67

3.55 ± 0.57

 < 0.001

2.99 ± 0.67

3.48 ± 0.56

 < 0.001

Lesion conspicuity score

3.70 ± 0.61

3.78 ± 0.55

0.046

3.72 ± 0.63

3.83 ± 0.38

0.107

3.61 ± 0.63

3.93 ± 0.26

0.001