Table 2 Optimization of the experimental variables affecting the performance of the NADH electrocatalytic sensor using mouse blood.

From: Disposable electrocatalytic sensor for whole blood NADH monitoring

Variable

Tested

Selected

Blood dilution factor

1:10, 1:20, 1:40, 1:80, 1:160

1:40

EDTA concentration, mM

0, 0.05, 0.5, 5, 50, 500

5

Applied potential, mV

500, 600, 700, 800

700