Table 1 Statistical parameters defining the correlation of the three partitions of the calibration dataset (see text for details) with the three blood parameters.

From: Non-invasive estimation of hemoglobin, bilirubin and oxygen saturation of neonates simultaneously using whole optical spectrum analysis at point of care

Amount of scattered data from the cumulative fitted curve

Statistical test

Parameters

Hb measurement

TSB measurement

SpO2 measurement

90%

Linear regression analysis

Regression Coefficient (r)

0.77

0.74

0.70

P value

 < 0.0001

 < 0.0001

 < 0.0001

Slope

0.9117

0.852

0.961

Intercept

1.1749

2.564

2.375

95%

Linear regression analysis

Regression Coefficient (r)

0.87

0.84

0.88

P value

 < 0.0001

 < 0.0001

 < 0.0001

Slope

0.917

0.94

1.0294

Intercept

1.211

0.844

2.098

98%

Linear regression analysis

Regression Coefficient (r)

0.98

0.98

0.98

P value

 < 0.0001

 < 0.0001

 < 0.0001

Slope

0.9548

1.006

0.977

Intercept

0.6686

0.372

1.775