Table 2 Insertion depth of intraocular lens injector systems into a clear corneal incision.

From: Experimental analysis of recommended corneal incision sizes in cataract surgery using 13 intraocular lens injector systems

Injector system

Manufacturer

Insertion depth (mm)

Mean

SD

95% CI

AuotonoMe

Alcon Inc

2.75

0.18

2.69–2.81

Ultrasert

Alcon Inc

2.62

0.19

2.54–2.70

MultiSert

Hoya Medical Singapore Ltd

5.53

0.37

5.41–5.65

Kowa original injector

Kowa Company Ltd

6.04

0.33

5.93–6.15

Accuject 1.6-1P

Medicel AG

4.54

0.29

4.38–4.70

Accuject 1.8-1P

Medicel AG

6.00

0.55

5.83–6.17

Accuject 2.2-1P

Medicel AG

6.06

0.43

5.92–6.20

Ergoject 2.2-TL

Medicel AG

5.50

0.45

5.25–5.74

Viscoject-Bio 2.2

Medicel AG

5.57

0.50

5.34–5.80

Ray Sert PLUS

Rayner Intraocular Lenses Ltd

6.33

0.63

6.17–6.48

RayOne

Rayner Intraocular Lenses Ltd

6.30

0.64

6.05–6.55

Bluemixs 180

Carl Zeiss Meditec AG

6.20

0.51

6.04–6.37

SkyJet

Carl Zeiss Meditec AG

6.64

1.23

5.96–7.33