Table 2 Comparison of Elecsys ECLIA N-antigen measurements in EDTA/Heparin plasma without vs with addition of the extraction buffer. (95% confidence intervals).

From: Sensitivity and performance of three novel quantitative assays of SARS-CoV-2 nucleoprotein in blood

Sample type

EDTA without vs with extraction buffer

Heparin without vs with extraction buffer

Number of samples

10

10

Slope

1.03 (0.96 to 1.1)

0.99 (0.86 to 1.1)

Y-intercept

− 0.046 (− 0.31 to 0.22)

0.14 (− 0.41 to 0.68)

Spearman r

 > 0.99

 > 0.99