Figure 2 | Scientific Reports

Figure 2

From: Infrared neural stimulation markedly enhances nerve functionality assessment during nerve monitoring

Figure 2

Infrared neural stimulation more readily detects crush injuries. (a) Normalized CMAP amplitudes resulting from ES in baseline, healthy, partially crush, and complete crush conditions. Black dashed line represents the amplitude damage threshold (50% decrease). (b) Normalized CMAP amplitudes resulting from INS in baseline, healthy, partially crush, and complete crush conditions. (c) Sensitivity and specificity for the amplitude-based IONM approach to crush injuries. (d) Normalized CMAP latencies resulting from ES in baseline, healthy, partial crush, and complete crush conditions. Black dashed line represents the latency damage threshold (10% increase). If no CMAP was evoked latency was set to 1.2 for ease of visual interpretation. (e) Normalized CMAP latencies resulting from INS in baseline, healthy, partial crush, and complete crush conditions. (f) Sensitivity and specificity for the latency-based IONM approach to crush injuries. Specificity in all cases was calculated using the ‘Healthy’ category of responses. All data is normalized to the mean baseline values for each individual nerve. \(n = 10\) nerves for all data sets (5 rats).

Back to article page