Table 2 TLC system suitability testing parameters for the determination of nirmatrelvir and ritonavir.
Parameters | Obtained value | Reference value23 | |
|---|---|---|---|
Nirmatrelvir | Ritonavir | ||
Retardation factor (Rf) | 0.14 | 0.41 | – |
Retention factor (K′) | 9.00 | 1.70 | 1–10 |
Tailing factor (T) | 1.50 | 1.10 | > 2 |
Resolution (Rs) | 4.13 | < 2 | |