Table 3 Similarity of ventricular depolarization metrics measured from 99-lead BSPM and 12-lead ECG.
From: Localization of the ventricular pacing site from BSPM and standard 12-lead ECG: a comparison study
Parameter | 12-lead ECG | BSPM | ICC | 95% CI | P value |
|---|---|---|---|---|---|
Intrinsic rhythm | |||||
TAT, ms | 169 ± 23 | 166 ± 24 | 0.94 | 0.84–0.98 | < 0.0001 |
LVAT, ms | 144 ± 20 | 143 ± 19 | 0.93 | 0.81–0.97 | < 0.0001 |
VEU, ms | 48 ± 12 | 42 ± 17 | 0.83 | 0.52–0.94 | < 0.0001 |
LV pacing | |||||
TAT, ms | 179 ± 26 | 180 ± 29 | 0.84 | 0.58–0.94 | < 0.0001 |
LVAT, ms | 156 ± 21 | 153 ± 21 | 0.83 | 0.58–0.95 | < 0.0001 |
VEU, ms | -50 ± 32 | -39 ± 35 | 0.85 | 0.60–0.94 | < 0.0001 |
RV pacing | |||||
TAT, ms | 177 ± 24 | 179 ± 21 | 0.95 | 0.86–0.98 | < 0.0001 |
LVAT, ms | 143 ± 24 | 140 ± 23 | 0.92 | 0.78–0.97 | < 0.0001 |
VEU, ms | 31 ± 16 | 30 ± 16 | 0.94 | 0.84–0.98 | < 0.0001 |