Table 3 Specs of the low-cost LF-FD-OCT and reported low-cost SD-OCT systems.

From: Rapid imaging and product screening with low-cost line-field Fourier domain optical coherence tomography

 

Low-cost

LF-FD-OCT

Low-cost

SD-OCT9

Low-cost

SD-OCT10

Low-cost

SD-OCT11

Light source

SLD

SLD

SLD

SLD

Centre wavelength

833 nm

840 nm

830 nm

830 nm

FWHM bandwidth

40 nm

50 nm

45 nm

42 nm

Power of illumination

1.8 mW

 ~ 1.3 mW

0.7 mW

0.68 mW

B-scan range

4.0 mm

Flexiblea

7.0 mm

6.6 mm

Axial resolution (in air)

8.3 μm

8.1 μm

7.0 μm

8.0 μm

Lateral resolution (in air)

11.0 μm

21.4 μm

17.6 μm

19.6 μm

Imaging depth (in air)

2.0 mm

2.7 mm

2.8 mm

2.7 mm

A-scan rate

100,000

A-scans/s

10,000

A-scans/s

8,800

A-scans/s

12,500

A-scans/s

Sensitivity

85 dB

98.89 dB

99.4 dB

104 dB

System cost

£6,122

 ~ £5,900

 ~ £5,800

 ~ £4,100