Table 2 Longitudinal chromatic aberration (LCA) of the studied IOLs at the three designed foci. The LCA is the difference between the nominal power measured at 480 nm and 644 nm. All reported values correspond to the IOL plane. Int. = intermediate.
From: Chromatic aberration and spectral dependency of extended-range-of-vision intraocular lens technology
Model | Trinova Pro C | PanOptix | Synergy | AT Lisa | Triumf |
|---|---|---|---|---|---|
Far [D] | 0.96 ± 0.00 | 0.92 ± 0.02 | − 0.37 ± 0.07 | 0.55 ± 0.01 | 0.10 ± 0.01 |
Int. [D] | 0.47 ± 0.00 | 0.41 ± 0.02 | − 1.37 ± 0.04 | 0.01 ± 0.03 | − 0.34 ± 0.05 |
Near [D] | 0.03 ± 0.07 | 0.04 ± 0.00 | − 0.99 ± 0.03 | − 0.46 ± 0.00 | − 0.85 ± 0.01 |