Table 2 The nominal dose delivered is 3 Gy.
From: New opto-electro-mechanical sensor for two-dimensions dosimetry based on radiochromic films
PMMA step height (mm) | Measured dose (Gy) in the EBT3 film | Monte Carlo simulation | |||
|---|---|---|---|---|---|
Scanner | dosiMOEMs (position #1) | dosiMOEMs (position #2) | Experimental dose ratio | Simulated dose ratio | |
1 | 2.9 ± 0.3 | – | 2.8 ± 0.3 | 1.07 ± 0.23 | 1.04 ± 0.02 |
2 | 2.9 ± 0.3 | 2.7 ± 0.3 | 2.7 ± 0.3 | 1.11 ± 0.23 | 1.09 ± 0.02 |
3 | 2.9 ± 0.3 | 2.6 ± 0.3 | 2.6 ± 0.3 | 1.15 ± 0.25 | 1.13 ± 0.02 |
4 | 2.8 ± 0.3 | 2.5 ± 0.3 | 2.6 ± 0.3 | 1.15 ± 0.25 | 1.17 ± 0.02 |
5 | 2.8 ± 0.3 | 2.5 ± 0.3 | 2.4 ± 0.3 | 1.20 ± 0.26 | 1.21 ± 0.02 |
10 | 2.32 ± 0.23 | 2.1 ± 0.2 | – | 1.43 ± 0.28 | 1.42 ± 0.02 |