Table 3 comparison between different analytical techniques for the determination of NH4+ in water.
Technique | Chemistry | Linear Range (µM) | LOD (µM) | References |
|---|---|---|---|---|
Modified SIAa (termed ABA) | OPAb-Sulfide with Formaldehyde | 0.005–25 | 0.001 | |
SFA -LWCCc | IPBd-Phenol | 0.005–10 | 5.5 nM | |
Programmable flow Spectrophotometric | pH-indicator bromothymol blue | 0.028–55.6 | 15 nM | |
LCe with fluorescence detection | OPA-Sulfide | 0.625–10 | 330 nM | |
SWV-electrochemistry | Ag-PAAQ-MWCNT-CPE | 5–100 | 0.03 | This work |