Table 2 Descriptive statistics for the comparison between measurements acquired on a Q-TOF by MS1 (DIA) and MS2 (DIA), or on a QQQ by MRM, for a six-point dilution series of a lipid extract from commercially available human plasma.

From: Q-RAI data-independent acquisition for lipidomic quantitative profiling

 

Injection Volume (μL)

0.25

0.5

1

2

4

8

DIA MS1

 Total Number of Features

106

  Median RSD (%)

3.80

2.35

1.82

1.09

1.08

0.92

 Number of Features with RSD > 30%

1 (0.94%)

0 (0%)

0 (0%)

3 (2.83%)

2 (1.89%)

1 (0.94%)

  Median R2

0.990

 Number of Features with R2 < 0.8

5 (4.72%)

DIA MS2

 Total Number of Quantified Fragments

651 out of 1007 (64.65% of all fragments)

  Median RSD (%)

15.15

11.02

8.27

6.06

5.42

4.74

 Number of Fragments with RSD > 30%

167 (25.65%)

108 (16.59%)

73 (11.21%)

61 (9.37%)

37 (5.68%)

58 (8.91%)

  Median R2

0.957

 Number of Fragments with R2 < 0.8

143 (21.97%)

MRM

 Total Number of Transitions

287

  Median RSD (%)

5.74

1.71

1.41

1.68

1.38

5.51

 Number of Transitions with RSD > 30%

0 (0%)

0 (0%)

0 (0%)

0 (0%)

0 (0%)

0 (0%)

  Median R2

0.997

 Number of Transitions with R2 < 0.8

0 (0%)