Table 5 Repeatability of the APW features.

From: Arterial and venous peripheral vascular assessment using wearable electro-resistive morphic sensors

APW feature

Intra-class correlation of measurements

95%-confidence interval

Rise Time (PPG)

0.747

0.609 < ICC < 0.851

Rise Time (HeMo)

0.479

0.282 < ICC < 0.662

Half Pulse Width (PPG)

0.861

0.774 < ICC < 0.921

Half Pulse Width (HeMo)

0.35

0.144 < ICC < 0.559

Three Quarters Pulse Width (PPG)

0.372

0.169 < ICC < 0.577

Three Quarters Pulse Width (HeMo)

0.188

− 0.01 < ICC < 0.414

Dicrotic Notch to Diastolic Peak Time (PPG)

0.667

0.504 < ICC < 0.798

Dicrotic Notch to Diastolic Peak Time (HeMo)

0.181

− 0.011 < ICC < 0.405

Systolic peak to diastolic peak time (PPG)

0.881

0.805 < ICC < 0.933

Systolic peak to diastolic peak time (HeMo)

0.209

0.008 < ICC < 0.436

Stiffness Index (PPG)

0.854

0.764 < ICC < 0.917

Stiffness Index (HeMo)

0.294

0.086 < ICC < 0.513

Augmentation Index (PPG)

0.926

0.876 < ICC < 0.959

Augmentation Index (HeMo)

0.324

0.116 < ICC < 0.538

Pulse Width (PPG)

0.879

0.802 < ICC < 0.932

Pulse Width (HeMo)

0.823

0.714 < ICC < 0.899