Table 2 Results of the recovery of different concentrations of FAdV-I from clinical samples.

From: A sandwich amperometric immunosensor for the detection of fowl adenovirus group I based on bimetallic Pt/Ag nanoparticle-functionalized multiwalled carbon nanotubes

No.

Initial FAdV-I concentration in the sample (EID50 mL−1)

Added amount of FAdV-I (EID50 mL−1)

Total found

Recovery rate (%) (n = 5)

Average (EID50 mL−1)

RSD (%) (n = 5)

1

107.61

100.00

211.49

3.72

101.86

2

275.42

200.00

487.24

2.85

102.48

3

427.25

400.00

813.75

3.54

98.37

4

531.74

500.00

1059.31

4.71

102.67

5

693.21

600.00

1350.27

2.63

104.41

6

721.35

700.00

1463.79

3.28

102.99