Table 1 Repeatability for CenterVue MAIA and Nidek MP-3 in all subfields and for all age-groups.

From: Inter and intradevice assessment of microperimetry testing in aging eyes

Group

Nb of Stimuli

MAIA (dB)

MP-3 (dB)

SD Difference

CoR

Upper LoA

Lower LoA

SD Difference

CoR

Upper LoA

Lower LoA

Macula overall

900

2.35

\(\pm\) 4.61

4.73

− 4.49

2.32

\(\pm\) 4.55

4.72

− 4.37

Subfields

 Foveal

180

2.66

\(\pm\) 5.21

5.16

− 5.28

2.36

\(\pm\) 4.62

4.71

− 4.53

 Parafoveal

240

2.19

\(\pm\) 4.29

4.48

− 4.12

2.42

\(\pm\) 4.74

4.04

− 5.44

 Perifoveal

400

2.27

\(\pm\) 4.45

4.54

− 4.36

2.1

\(\pm\) 4.12

4.14

− 4.09

Superior quadrant

160

2.149

\(\pm\) 4.21

4.34

− 4.08

2.2

\(\pm\) 4.31

4.42

− 4.20

Nasal quadrant

160

2.002

\(\pm\) 3.92

3.91

− 3.93

2.1

\(\pm\) 4.12

4.05

− 4.19

Inferior quadrant

160

2.68

\(\pm\) 5.25

5.50

− 5.00

2.17

\(\pm\) 4.25

4.26

− 4.24

Temporal quadrant

160

2.08

\(\pm\) 4.08

4.10

− 3.96

2.44

\(\pm\) 4.78

4.64

− 4.92

Age-group in years

 Under 60

225

1.73

\(\pm\) 3.39

3.56

− 3.22

1.91

\(\pm\) 3.7

3.5

− 3.98

 Between 60 and 80

450

2.37

\(\pm\) 4.65

5.0

− 4.3

2.3

\(\pm\) 4.51

4.76

− 4.26

 Over 80

225

2.76

\(\pm\) 5.40

5.02

− 5.79

2.66

\(\pm\) 5.21

4.91

− 5.51