Table 2 The time average over the cardiac cycle of the standard deviation as a measure of beat-to-beat variation for the recorded left atrial pressure, left ventricular pressure, mitral valve flow, pulmonary venous flow, and aortic flow for the artificial left atrium number 1, 5, 10, and 11.

From: Investigating the importance of left atrial compliance on fluid dynamics in a novel mock circulatory loop

LA

LA pressure

MV flow

PV flow

LV pressure

AV flow

1

0.28

3.01

0.95

0.94

4.10

5

0.23

2.30

0.75

0.73

2.61

10

0.14

2.32

0.72

0.61

4.13

11

0.15

2.83

0.67

0.80

4.08

  1. AV aortic valve, LA left atrium, LV left ventricle, MV mitral valve, PV pulmonary vein.