Table 2 The BIC unit's recording capabilities and general properties.

From: Benchmarking signal quality and spatiotemporal distribution of interictal spikes in prolonged human iEEG recordings using CorTec wireless brain interchange

System recording capabilities

 Recording channels

32

Sampling rate

1kHz

 ADC resolution

16 Bits

Frequency band-pass

~ 2 to 325 Hz

 Amplification gains

750× (57.5 dB), 375× (51.5 dB), 188× (45.5 dB), 95× (39.5dB)

 Max. input voltage

4–30 \({{\text{mV}}}_{{\text{pp}}}\) (dependent on the amplification gain)

System general properties

 Implant dimensions (mm3)

67 × 38 × 7

Power supply

Wireless

 Stimulation frequency range

Single pulse-200 Hz

Max. stimulation pulse current

6.12 mA

 Stimulation pulse width range

10–2500 µs

Charge balancing counter pulse

4× pulse width

 Stimulation pulse shape

Asymmetric rectangular and biphasic fashion

 Communication

Bi-directional, radio frequency (2.45 GHz), 5 Mbit/s

 Powering

Inductive 120–140 kHz Power regulated on implant demand

 BIC reference

Any electrode contacts or group of electrode contacts