Table 3 Intraclass correlation coefficient (95% confidence interval) values of retinal thickness measurements from four different optical coherence tomography devices.

From: Comparison of retinal thickness measurements among four different optical coherence tomography devices

 

PLEX Elite and DRI OCT-1 Atlantis

PLEX Elite and Cirrus 5000 HD-OCT

PLEX Elite and Spectralis

DRI OCT-1 Atlantis and Cirrus 5000 HD-OCT

DRI OCT-1 Atlantis and Spectralis

Cirrus 5000 HD-OCT and Spectralis

Center

0.876 (0.694–0.950)

0.852 (0.635–0.940)

0.894 (0.740–0.957)

0.902 (0.758–0.960)

0.964 (0.912–0.986)

0.928 (0.822–0.971)

Inner

 Superior

0.830 (0.581–0.931)

0.741 (0.362–0.895)

0.826 (0.571–0.929)

0.479 (–0.283–0.789)

0.960 (0.902–0.9984)

0.547 (–0.116–0.816)

 Temporal

0.826 (0.571–0.929)

0.799 (0.505–0.919)

0.766 (0.423–0.905)

0.824 (0.565–0.928)

0.936 (0.843–0.974)

0.744 (0.369–0.896)

 Inferior

0.790 (0.482–0.915)

0.694 (0.247–0.876)

0.825 (0.569–0.929)

0.864 (0.665–0.945)

0.956 (0.891–0.982)

0.875 (0.692–0.949)

 Nasal

0.921 (0.804–0.968)

0.872 (0.686–0.948)

0.935 (0.839–0.974)

0.803 (0.514–0.920)

0.911 (0.780–0.964)

0.749 (0.382–0.898)

Outer

 Superior

0.871 (0.682–0.948)

0.734 (0.343–0.892)

0.858 (0.649–0.942)

0.873 (0.688–0.949)

0.964 (0.912–0.985)

0.877 (0.697–0.950)

 Temporal

0.690 (0.237–0.874)

0.744 (0.370–0.896)

0.496 (–0.243–0.795)

0.930 (0.828–0.972)

0.528 (–0.162–0.809)

0.370 (–0.552–0.744)

 Inferior

0.867 (0.673–0.946)

0.900 (0.755–0.960)

0.865 (0.667–0.945)

0.928 (0.822–0.971)

0.977 (0.942–0.990)

0.905 (0.765–0.961)

 Nasal

0.914 (0.789–0.965)

0.887 (0.723–0.954)

0.922 (0.808–0.968)

0.883 (0.712–0.953)

0.964 (0.912–0.986)

0.899 (0.752–0.959)