Table 4 The chemistry analyzers sample demonstrated the values P, S, and D according to five experts’ ratings.
From: An integrative approach to medical laboratory equipment risk management
| Â | Â | Experts severity | Experts probability | Experts detection | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Device No. | Criterion | #1 | #2 | #3 | #4 | #5 | #1 | #2 | #3 | #4 | #5 | #1 | #2 | #3 | #4 | #5 |
1 | C1 | 5 | 4 | 5 | 4 | 5 | 5 | 4 | 5 | 4 | 5 | 2 | 1 | 3 | 2 | 2 |
7 | 4 | 5 | 3 | 4 | 4 | 3 | 2 | 4 | 3 | 3 | 2 | 1 | 3 | 2 | 2 | |
25 | 3 | 2 | 4 | 3 | 3 | 3 | 2 | 4 | 3 | 3 | 2 | 2 | 1 | 2 | 3 | |
50 | 5 | 4 | 5 | 4 | 5 | 5 | 4 | 5 | 4 | 5 | 3 | 3 | 3 | 2 | 4 | |
1 | C2 | 5 | 5 | 4 | 4 | 5 | 5 | 5 | 4 | 4 | 5 | 1 | 1 | 1 | 1 | 2 |
7 | 2 | 2 | 1 | 2 | 3 | 3 | 3 | 3 | 2 | 4 | 2 | 2 | 1 | 2 | 3 | |
25 | 4 | 5 | 3 | 4 | 4 | 4 | 5 | 3 | 4 | 4 | 1 | 1 | 1 | 1 | 2 | |
50 | 5 | 4 | 5 | 4 | 5 | 5 | 4 | 5 | 4 | 5 | 2 | 1 | 3 | 2 | 2 | |
1 | C3 | 4 | 5 | 3 | 4 | 4 | 4 | 5 | 3 | 4 | 4 | 2 | 2 | 1 | 2 | 3 |
7 | 4 | 5 | 3 | 4 | 4 | 4 | 5 | 3 | 4 | 4 | 2 | 2 | 1 | 2 | 3 | |
25 | 3 | 3 | 3 | 2 | 4 | 3 | 3 | 3 | 2 | 4 | 2 | 2 | 1 | 2 | 3 | |
50 | 4 | 5 | 3 | 4 | 4 | 4 | 5 | 3 | 4 | 4 | 2 | 1 | 3 | 2 | 2 | |
1 | C4 | 5 | 4 | 5 | 4 | 5 | 4 | 3 | 4 | 4 | 5 | 2 | 2 | 1 | 2 | 3 |
7 | 3 | 3 | 3 | 2 | 4 | 3 | 3 | 3 | 2 | 4 | 3 | 2 | 4 | 3 | 3 | |
25 | 5 | 4 | 5 | 4 | 5 | 5 | 4 | 5 | 4 | 5 | 2 | 1 | 3 | 2 | 2 | |
50 | 4 | 5 | 3 | 4 | 4 | 4 | 5 | 3 | 4 | 4 | 2 | 2 | 1 | 2 | 3 | |
1 | C5 | 3 | 3 | 3 | 2 | 4 | 3 | 3 | 3 | 2 | 4 | 2 | 2 | 1 | 2 | 3 |
7 | 5 | 4 | 5 | 4 | 5 | 5 | 4 | 5 | 4 | 5 | 1 | 1 | 1 | 1 | 2 | |
25 | 4 | 3 | 4 | 4 | 5 | 4 | 3 | 4 | 4 | 5 | 2 | 1 | 3 | 2 | 2 | |
50 | 5 | 4 | 5 | 4 | 5 | 5 | 4 | 5 | 4 | 5 | 1 | 1 | 1 | 1 | 2 | |