Table 3 Contribution of primary and scattered photons in the fluence and average energy of 6 MV photons at different depths inside the water for open and grid fields of 10 × 10 and 20 × 20 cm2. The relative uncertainty in fluence calculations was kept within 1%.

From: Beam quality and the mystery behind the lower percentage depth dose in grid radiation therapy

Depth

Open field

Grid field

Brass grid

Cerrobend grid

Primary

Secondary

Total

Primary

Secondary

Total

Primary

Secondary

Total

 

Photon fluence (108/cm2/MU)

10 × 10 cm2 treatment field

 1.5 cm

12.89

4.09

16.98

11.46

1.29

12.75

11.43

1.16

12.59

 5 cm

9.74

5.67

15.41

8.67

1.63

10.30

8.65

1.47

10.12

 10 cm

6.62

5.55

12.17

5.91

1.55

7.47

5.90

1.39

7.29

Photon average energy (MeV)

 1.5 cm

1.83

0.35

1.47

1.90

0.59

1.77

1.90

0.57

1.78

 5 cm

1.92

0.47

1.38

1.99

0.63

1.78

1.99

0.62

1.79

 10 cm

2.04

0.55

1.36

2.12

0.69

1.82

2.12

0.67

1.84

Photon fluence (108/cm2/MU)

20 × 20 cm2 treatment field

 1.5 cm

13.22

6.94

20.17

11.50

2.31

13.81

11.45

1.93

13.39

 5 cm

10.02

9.80

19.82

8.71

2.92

11.63

8.68

2.50

11.18

 10 cm

6.86

10.36

17.22

5.95

2.96

8.91

5.93

2.56

8.49

Photon average energy (MeV)

 1.5 cm

1.81

0.28

1.28

1.89

0.57

1.67

1.90

0.52

1.70

 5 cm

1.89

0.35

1.13

1.99

0.54

1.62

1.99

0.51

1.66

 10 cm

2.00

0.40

1.03

2.11

0.54

1.59

2.11

0.52

1.63