Table 3 Linearity data of analytical methods.
Parameter | HPLC method | Spectrophotometric method |
|---|---|---|
Range of concentration [μg mL−1] [n = 6] | 5–30 | |
The slope of the regression equation | 50.0930 | 0.0435 |
The intercept of the regression equation | 2.8095 | − 0.0083 |
Correlation coefficient | 0.9997 | 0.9995 |
Limit of detection [μg mL−1] | 0.70 | 0.90 |
Limit of quantification [μg mL−1] | 2.00 | 2.80 |
Recovery % [n = 3] | 99.52–100.58 | 98.99–101.07 |