Table 4 Results of the precision of the methods.

From: A green approach for metoclopramide quantification in pharmaceutical products: new HPLC and spectrophotometric methods

Precision

HPLC method

Spectrophotometric method

Retention time, min.

Peak area

Assay %

Absorbance

Assay %

Intraday

Average

3.038

1253.68

100.00

1.070

100.00

SD

0.001

0.781

0.062

0.001

0.108

RSD %

0.033

0.062

0.062

0.108

0.108

Interday

Average

3.037

1252.67

100.00

1.074

100.01

SD

0.002

1.527

0.122

0.002

0.135

RSD %

0.079

0.122

0.122

0.135

0.135