Table 2 UEQ-S results (n = 35).
From: Design and validation of a wearable dynamometry system for knee extension-flexion torque measurement
IKD (Biodex) | WDS | CIM [95% CI] | p-value | |
|---|---|---|---|---|
Supportive | 1.46 ± 1.20 | 1.94 ± 0.80 | 0.49 [0.24, 0.73] | 0.051 |
Easy | 0.31 ± 1.98 | 2.14 ± 1.03 | 1.83 [1.45, 2.21] | < 0.001 |
Efficient | 0.31 ± 1.75 | 1.94 ± 1.00 | 1.63 [1.29, 1.97] | < 0.001 |
Clear | 0.46 ± 1.84 | 1.91 ± 0.95 | 1.46 [1.11, 1.81] | < 0.001 |
Exciting | 1.17 ± 1.10 | 1.46 ± 0.98 | 0.29 [0.04, 0.53] | 0.255 |
Interesting | 1.14 ± 1.06 | 1.57 ± 1.12 | 0.43 [0.17, 0.69] | 0.105 |
Inventive | 0.31 ± 1.53 | 1.97 ± 0.92 | 1.66 [1.36, 1.96] | < 0.001 |
Leading edge | 0.34 ± 1.51 | 1.74 ± 0.89 | 1.40 [1.10, 1.70] | < 0.001 |
Total score | 5.51 ± 9.95 | 14.69 ± 6.30 | 9.17 [7.19, 11.20] | < 0.001 |