Table 2 UEQ-S results (n = 35).

From: Design and validation of a wearable dynamometry system for knee extension-flexion torque measurement

 

IKD (Biodex)

WDS

CIM [95% CI]

p-value

Supportive

1.46 ± 1.20

1.94 ± 0.80

0.49 [0.24, 0.73]

0.051

Easy

0.31 ± 1.98

2.14 ± 1.03

1.83 [1.45, 2.21]

 < 0.001

Efficient

0.31 ± 1.75

1.94 ± 1.00

1.63 [1.29, 1.97]

 < 0.001

Clear

0.46 ± 1.84

1.91 ± 0.95

1.46 [1.11, 1.81]

 < 0.001

Exciting

1.17 ± 1.10

1.46 ± 0.98

0.29 [0.04, 0.53]

0.255

Interesting

1.14 ± 1.06

1.57 ± 1.12

0.43 [0.17, 0.69]

0.105

Inventive

0.31 ± 1.53

1.97 ± 0.92

1.66 [1.36, 1.96]

 < 0.001

Leading edge

0.34 ± 1.51

1.74 ± 0.89

1.40 [1.10, 1.70]

 < 0.001

Total score

5.51 ± 9.95

14.69 ± 6.30

9.17 [7.19, 11.20]

 < 0.001

  1. CIM change in mean, CI confidence interval, WDS wearable dynamometry system, IKD isokinetic dynamometry.
  2. Significant values are in bold.