Figure 7
From: A high-resolution large-area detector for quality assurance in radiotherapy

Bragg curves. Increasing energy deposition with penetration depth, with a maximum at the end of the range (151 mm) followed by a sharp decrease. The curves are normalized to unity in the plateau region (50 mm depth in water). (a) Bragg curve measured with the GEM + TFT (
) and PTW Peakfinder (
) with 147.72 MeV protons. (b) Bragg curve measured with the GEM + TFT (
) and PTW Peakfinder (
) with 207.97 MeV/n carbon ions.