Table 3 Regression and validation parameters of the proposed HPLC method for the determination of pure DIP, HBB and AMP in bovine urine.
Parameters | Standards | ||
|---|---|---|---|
DIP | HBB | AMP | |
Range µg/mL | 15–75 | 2.5–60 | 2.5–60 |
Slope | 1.1178 | 0.7914 | 0.1939 |
intercept | − 11.357 | 0.2956 | 0.2237 |
Correlation coefficient | 0.9999 | 0.9998 | 0.9999 |
Accuracy amean ± SD | 100.30 ± 0.23 | 100.27 ± 0.41 | 99.46 ± 0.48 |
Precision (RSD%) bRepeatability cIntermediate precision | 0.332 0.124 | 0.446 0.288 | 0.426 0.600 |
dLLOQ µg/mL | 15 | 2.5 | 2.5 |
eLQC µg/mL | 45 | 7.5 | 7.5 |
fMQC µg/mL | 55 | 25 | 25 |
gHQC µg/mL | 60 | 50 | 50 |