Fig. 6 | Scientific Reports

Fig. 6

From: Mechanism behind reduced resection ability of bent snares during cold snare polypectomy

Fig. 6

Resection completion rate of cold snare polypectomy using snares bent one time or five times. To evaluate the performance of the bent snares, cold snare polypectomy was performed for a 5-mm-diameter pseudo-lesion using a porcine rectum, and the resection completion rate was calculated. The resection completion rate was defined as the percentage of pseudo-lesions that could be resected with a snare without becoming stuck. Pathological evaluations were not included because pseudo-lesions were targeted. Data are expressed as means ± standard errors. *P < 0.05 (versus no bending). See also Supplementary Table S1 for detailed data.