Table 3 MRI Acquisition parameters in Phantom and Clinical Study.
Sequence | Phantom study | Clinical study |
|---|---|---|
Acquisition mode | Cartesian | Cartesian |
Shot mode | Single-shot | Single-shot |
Freq FOV (cm) / phase FOV | 256 × 256 | 380 × 300 |
Slice thickness / spacing (mm) | 1 / 0 | 3 / 0 |
Number of slice | 1 | 2 |
Cross sectional image | Coronal | Sagittal |
Voxel size (reconstructed) | 1.0 × 1.0 × 1.0 | 1.0 × 1.0 × 3.0 |
TR (ms) | 2.7 | 3.1 |
TE (ms) | 1.23 | 1.46 |
Flip angle (degree) | 5/10/15/20/25/30/35/40/45/50 | 30 |
Bandwidth (Hz) | 868.1 | 1083.5 |
Fat suppression | SPIR | SPIR |
TFE factor | 32 | 40 |
Shot interval | Shortest | Shortest |
TFE prepulse / pulse type | Saturate/Composite | Saturate / Composite |
TFE prepulse delay (ms) | 50/100/200/300/400 | 150 |
TFE shot duration (ms) | 246.8 | 285 |
Accelarated factor | CS 4 | SENSE 4 / CS 4 |
Denoising level | Strong | No / Strong |