Table 5 Comparisons of intra-observer and inter-observer absolute differences in FNPO assessment.

From: Enhancing repeatability of follicle counting with deep learning reconstruction high-resolution MRI in PCOS patients

Parameter

SSFSE-DL

SSFSE-C

PROPELLER

SSFSE-DL versus SSFSE-C

SSFSE-DL versus PROPELLER

Intra-observer difference

1.796 ± 0.930 (1.513–2.078)

2.455 ± 1.266 (2.070–2.839)

3.046 ± 1.311 (2.647–3.444)

P = 0.002

P < 0.001

Inter-observer difference

2.023 ± 0.821 (1.773–2.272)

3.182 ± 1.206 (2.815–3.549)*

4.000 ± 1.643 (3.501–4.499)*

P < 0.001

P < 0.001

  1. Absolute differences are presented as mean ± standard deviation, with 95% confidence intervals in parentheses. Comparisons were conducted using paired t tests, and results with P-values below 0.05 were emphasized in bold. FNPO follicle number per ovary, SSFSE-C single-shot fast spin-echo acquisition using conventional reconstruction, SSFSE-DL single-shot fast spin-echo acquisition using deep learning reconstruction, PROPELLER periodically rotated overlapping parallel lines with enhanced reconstruction; *Statistically significant between the intra-observer and inter-observer absolute differences. Significant values are in bold.