Table 6 Comparison of oxygen saturation estimation methods.
From: Continuous non-contact vital sign monitoring of neonates in intensive care units using RGB-D cameras
Method | \(\downarrow\)MAE (%) | \(\downarrow\)MSE (%) | \(\uparrow\)CP\(\pm 3\%\) (%) | \(\uparrow\)CP\(\pm 6\%\) (%) |
|---|---|---|---|---|
Red/Infrared | 3.37 | 15.89 | 57.45 | 84.15 |
Red/Blue | 8.93 | 123.11 | 20.45 | 41.14 |
YCgCr | 3.27 | 22.53 | 64.23 | 74.23 |
Calibration free | 9.49 | 190.34 | 31.65 | 47.13 |