Table 1 Demographic data of participants. MTSL – maximum tolerable stimulation level, Y – years, M – months, f – female, m – male, r – right, l – left, C – cochleostomy, R – round window.

From: A case series shows a relation between intracochlear electric field distribution and vestibular co-stimulation with cochlear implants

Subject ID

Implant type

Duration of use

Age in Y

Sex

Test side

MTSL

Threshold of e-VEMP

Insertion

e-VEMP group

 eVEMP_ 05

CI24R

15 Y

55

f

r

220 CL

180 CL

C

 eVEMP_ 06

CI24RE

7 Y

59

f

r

200 CL

190 CL

R

 eVEMP_ 10

CI24RE

12 Y

56

f

r

235 CL

185 CL

C

 eVEMP_ 17

CI632

4 M

49

f

r

220 CL

220 CL

R

 eVEMP_ 19

CI512

6 Y

58

m

l

220 CL

220 CL

R

 eVEMP_ 21

CI632

1 Y

35

m

l

190 CL

175 CL

R

 eVEMP_ 23

CI24RE

8 Y

24

f

r

220 CL

200 CL

R

 eVEMP_ 24

CI512

5 Y

63

f

r

215 CL

215 CL

R

 eVEMP_ 25

CI632

1 Y

23

f

r

240 CL

200 CL

R

 eVEMP_ 26

CI24RE

13 Y

63

f

r

190 CL

190 CL

C

 Average e-VEMP group

 

6.8 +/- 5.2

48.5 +/- 15.5

  

215 +/- 17

200 +/- 15

 

no e-VEMP group

 eVEMP_ 01

CI512

8 Y

59

m

r

225 CL

R

 eVEMP_ 02

CI512

3 Y

42

f

l

200 CL

R

 eVEMP_03

CI24RE(CA)

7 Y

60

m

l

195 CL

R

 eVEMP_04

CI512

3 Y

61

f

r

210 CL

R

 eVEMP_07

CI632

2 M

36

m

l

215 CL

R

 eVEMP_09

CI532

1 Y

60

f

l

145 CL

R

 eVEMP_11

CI532

2 Y

47

f

l

215 CL

R

 eVEMP_12

CI532

2 Y

60

f

l

205 CL

R

 eVEMP_14

CI532

3 Y

43

f

l

205 CL

R

 eVEMP_15

CI632

1 Y

55

f

r

190 CL

R

 eVEMP_18

CI632

1 Y

60

f

l

185 CL

R

 eVEMP_20

CI512

4 Y

35

m

r

185 CL

C

 eVEMP_22

CI632

1 Y

58

f

r

175 CL

R

 eVEMP_16

CI532

2 Y

63

f

l

210 CL

R

 eVEMP_27

CI532

2 Y

47

m

r

225 CL

R

 Average

 no e-VEMP group

 

2.7 +/- 2.2

52.4 +/-9.7

  

199 +/- 21

  

 Average all

 

4.3 +/- 4.2

50.8 +/- 12.2

  

205 +/- 21