Table 2 Expert judgments for each basic event.

From: Improving maintenance efficiency and controlling costs in healthcare institutions through advanced analytical method

Symbol

Events

Symbol

Events

Symbol

Events

BE1

Power switch key malfunction

BE15

ECG signal processing module failure

BE29

Oxygen Clip Electrode Failure

BE 2

Main control board failure

BE16

Faulty connection between the ECG signal acquisition lead wire and the ECG signal measurement module

BE30

Torn sphygmomanometer cuff

BE 3

Power switch to keypad connection failure

BE17

Faulty connection between the ECG signal measurement module and the main control board

BE31

Amplifier Failure

BE 4

Battery failure

BE18

SPO2 signal measurement module failure

BE32

Alarm system failure

BE 5

Faulty connection between the battery interface board and the power board

BE19

Faulty connection between the SPO2 signal acquisition extension cable and SPO2 signal measurement module

T

Multi-parameter monitor failure

BE 6

Power control failure

BE20

Faulty connection between the SPO2 signal measurement module and the main control board

M1

Power-on failure

BE7

Faulty connection between the keypad and the main control board

BE21

NIBP signal acquisition air leak in the airway

M2

Power supply module failure

BE8

Faulty connection between the power board to main control board failure

BE22

Air pump failure

M3

Display failure

BE9

Display malfunction

BE23

Faulty connection between the NIBP signal acquisition tube and NIBP signal measurementmodule

M4

EEG signal module failure

BE10

Backlight failure

BE24

NIBP signal acquisition guide tube bent and blocked

M5

SPO2 signal module failure

BE11

High-voltage board failure

BE25

Faulty connection between the NIBP signal measurement module and main control board

M6

NIBP signal module failure

BE12

Faulty connection between the Display screen and main control board

BE26

Solenoid valve blocked

M7

Solenoid valve failure

BE13

Faulty connection between the display screen and backlight board

BE27

Circuit board failure

M8

External sensor failure

BE14

Digital video interface failure.

BE28

ECG Electrode Failure