Table 3 Mean values and deviations of differences in measured values between the NCVS and Ref.

From: A noncontact vital sign sensor demonstrating a strong correlation with an electrocardiogram electrode and a CO2 sensor

 

Mean differences between

the NCVS and Ref.

Deviation

All patients

 HR

− 5.3 bpm

12.9 bpm

 RR

− 1.0 breaths/min

3.1 breaths/min

Adults

 HR

− 2.3 bpm

8.6 bpm

 RR

− 0.6 breaths/min

1.5 breaths/min

Children

 HR

− 8.1 bpm

15.9 bpm

 RR

− 1.5 breaths/min

4.1 breaths/min

  1. NCVS, noncontact vital sign sensor; Ref, reference sensor; HR, heart rate; RR, respiratory rate.