Table 6 Factors associated with high errors with MTB test

From: Pilot implementation result of frequency, distribution, and factors associated with unsuccessful tests and error rates at Truenat sites in Nigeria

Variable

MTB errors (n = 38)

χ2 p value

High (≥ 3%) N (%)

Low (< 3%) N (%)

Age of respondents

   

< 38 years

6 (28.6)

15 (71.4)

0.197 0.658

≥ 38 years

6 (35.3)

11 (64.7)

 

Gender

   

Male

8 (29.6)

19 (70.4)

0.164 0.685

Female

4 (36.4)

7 (63.6)

 

Type of health facility

   

Public

9 (31.0)

20 (69.0)

0.017 0.897

Private

3 (33.3)

6 (66.7)

 

Facility is a DOTS center

   

Yes

9 (29.0)

22 (71.0)

FT 0.656

No

3 (42.9)

4 (57.1)

 

Facility is a microscopy center

   

Yes

6 (31.6)

13 (68.4)

FT 1.0

No

6 (31.6)

13 (68.4)

 

Previous experience with molecular diagnostic method

   

Yes

7 (33.3)

14 (66.7)

0.067 0.796

No

5 (29.4)

12 (70.6)

 

Laboratory staff trained by Molbio

   

Yes

10 (29.4)

24 (70.6)

FT 0.577

No

2 (50.0)

2 (50.0)

 

No of years of experience

   

< 10 years

7 (35.0)

13 (65.0)

0.229 0.632

≥ 10 years

5 (27.8)

13 (72.2)

 

No of years in TB program

   

≤ 5 years

4 (22.2)

14 (77.8)

1.386 0.2398

> 5 years

8 (40.0)

12 (60.0)

 

SOP for conducting Truenat visibly pasted on wall

   

Yes

11 (31.4)

24 (68.6)

FT 1.0

No

1 (33.3)

2 (66.7)

 

SOP for sample preparation for Truenat DNA extraction visibly pasted

   

Yes

9 (28.1)

23 (71.9)

FT 0.357

No

3 (50.0)

3 (50.0)

 

Preventive maintenance log form visibly kept and recorded

   

Yes

10 (32.3)

21 (67.7)

FT 1.0

No

2 (28.6)

5 (71.4)

 

Flushing protocol conducted and reported in form at least once monthly

   

Yes

10 (31.3)

22 (68.7)

FT 1.0

No

2 (33.3)

4 (66.7)

 

Slider glass change conducted and reported at least once monthly in past three months

   

Yes

11 (33.3)

22 (66.7)

FT 1.0

No

1 (20.0)

4 (80.0)

 
  1. FT Fisher’s exact test