Table 3 Comparative diagnostic performance of the research assay and InSignia assay in detecting viral infection (n = 38).
Metric | Research assay | InSignia assay |
|---|---|---|
Sensitivity [%, (95% CI)] | 75.86 (56.46–89.70) | 93.10 (77.23–99.15) |
Specificity [%, (95% CI)] | 100 (66.37–100) | 100 (66.37–100) |
Positive Predictive Value [%, (95% CI)] | 100 (84.56–100) | 100 (87.23–100) |
Negative Predictive Value [%, (95% CI)] | 56.25 (40.28–71.02) | 81.82 (54.16–94.49) |
Accuracy [%, (95% CI)] | 81.58 (65.67–92.26) | 94.74 (82.25–99.36) |