Table 2 MRE of the modeled complex impedance and modeled absolute value and MAE of the modeled phase angle referenced to the experimental data.

From: Electrical equivalent circuit for analyzing the effect of signal shape on power distribution in cochlear implant electrodes and surrounding tissue

 

SE 1 vs. SE 2

SE 1 vs. SE 6

SE 1 vs. SE 11

SE 1 vs. SE 16

Average (SE 1 vs. SE 2 to SE 16)

MRE of absolute value \(\overline{e}_{Z}\) (%)

4.19 ± 3.42

5.06 ± 6.65

6.32 ± 5.01

5.02 ± 3.51

5.40 ± 4.46

MAE of phase angle \(\overline{e}_{\varphi }\) (°)

2.30 ± 1.61

2.45 ± 2.39

2.24 ± 1.39

2.63 ± 2.19

2.39 ± 1.70

MRE of complex impedance \(\overline{e}_{{\underline {Z} }}\) (%)

6.48 ± 3.41

7.68 ± 6.71

8.03 ± 4.79

7.45 ± 4.41

7.56 ± 4.39