Fig. 1

From: Clinical performance validation of an integrated cartridge based bedside blood gas analyzer system in the acute care setting

Fig. 1

Bland-Altman Plots Demonstrating Agreement Between the EG and ABL. Bland-Altman plots illustrating the agreement between the EG and ABL for the ten clinical parameters: (a) pH, (b) pCO2, (c) pO2, (d) K+, (e) Na+, (f) iCa2+, (g) Cl, (h) Lac, (i) Glu, and (j) Hct. The plots display the difference between the two systems as a percentage of the average measurement across a range of values. The middle dashed line represents the mean difference, while the upper and lower dashed lines represent the limits of agreement (± 1.96 standard deviations). The mean difference and limits of agreement for each parameter are indicated on the respective plots.