Fig. 2 | Scientific Reports

Fig. 2

From: Clinical performance validation of an integrated cartridge based bedside blood gas analyzer system in the acute care setting

Fig. 2

Passing-Bablok Regression Analysis for the Agreement Between the EG and ABL90. Passing - Bablok regression analysis for each of the ten clinical parameters: (a) pH, (b) pCO2, (c) pO2, (d) K+, (e) Na+, (f) iCa2+, (g) Cl, (h) Lac, (i) Glu, and (j) Hct. The regression lines represent the relationship between the EG and ABL90, with the 95% confidence intervals shown as shaded areas. The equation for each regression line, along with the sample size (n = 216), is displayed in the upper-left corner of each plot.

Back to article page